Electrolytic Respiration-Based Biodegradation Tester

Electrolytic Respiration-Based Biodegradation Tester

Brand: Thmorgan
Model: T2400C

I. Product Applications

1. Can be used for biobased biodegradability testing, meeting the requirements of GB/T.   19276-1;

2. Suitable for seawater-based biodegradability testing, GB/T   40611-2021;

3. Applicable to potential biodegradation tests of materials under the action of specific microorganisms, GB/T 19275-2003;

4. Suitable for rapid bio-degradability testing of chemicals, per GB/T 21801, GB/T 21802, and GB/T 21803;

5. Meets international standards such as OECD301C, OECD301F, OECD302C, EC-partc4, and ISO5815-1:2003;

6. Meets international standards including OECD, EEC, ASTM, BBA, ISO, and MITI testing and research;

7. Can be used for the determination of BOD3, BOD5, BOD7, BOD28, BOD42, and more;

8. Can be used for biodegradation testing in the activated sludge process;

9. Suitable for microbial respiration measurement;

II. Product Features

1. Determine the actual oxygen consumption of the sample by measuring charge movement;

2. No need to dilute the sample: The maximum detection limit can reach 10,000 mg.

3. Continuous electrolytic oxygen generation: Each bottle of electrolyte can provide several years of electrolytic oxygen; simply add fresh electrolyte once the current electrolyte is depleted to continue use.

4. Precise temperature control: Air-bath temperature control with a set temperature range of 5–90℃ and a temperature adjustment accuracy of ±0.1℃.

5. Number of channels: Each instrument includes 24 detection channels;

6. Magnetic stirring system: Provides long-term stirring;

7. Data output system: Processed by dedicated software on a computer, with an LCD display;

III. Product Principle

Microorganisms consume the oxygen in the culture flask, producing carbon dioxide. The calcium oxide absorbs this carbon dioxide, causing the pressure inside the culture flask to decrease. This pressure difference between the two sides of the pressure gauge triggers the electrolytic circuit to close. As a result, the copper sulfate solution in the electrolytic cell undergoes electrolysis, generating oxygen that is replenished into the culture flask, gradually restoring the flask’s pressure to its initial level. The amount of electric charge passed into the electrolytic cell is directly proportional to the amount of oxygen produced; a coulometer accurately monitors the electric charge flowing into the electrolytic cell. Thus, the oxygen consumption can be calculated, and the ratio of actual oxygen consumption to theoretical oxygen consumption represents the biodegradation rate of the test sample.

 

IV. Technical Parameters

1. Working conditions:

1.1 Power supply voltage: AC 220V ±10%, 50Hz, single-phase;

1.2 Operating temperature: 15–30℃;

1.3 Relative humidity: ≤80%;

2. Oxygen supply system:

2.1 Oxygen Supply Method: Closed-loop copper sulfate electrolysis oxygen supply;

2.2 Internal liquid in the electrolytic cell: saturated copper sulfate solution;

2.3 Opening and closing of electrolysis: Air pressure on both sides of the U-tube;

2.4 Electrolytic accuracy: Sensitivity 9.806 Pa; internally filled with a saturated copper sulfate solution.

2.5 Electrodes: Platinum (positive electrode), copper (negative electrode);

2.6 Electrolytic bottle material: High-borosilicate hard glass;

3. Temperature control system:

3.1 Temperature control method: Air bath;

3.2 Set temperature range: 5–90℃, with a step size of 0.1℃;

3.3 Temperature accuracy: ≤ ±0.1℃;

4. Reaction system:

4.1 Reaction method: Liquid-phase reaction;

4.2 Reaction System: Closed-loop electrolytic reaction;

4.3 Stirring method: 24-channel magnetic stirring;

4.4 Culture flask volume: 500 ml;

4.5 Decarbonization System: None;

4.6 Condensation System: None;

5. Detection System:

5.1 Detection Principle: The coulometric method detects charge movement, with four charges corresponding to one oxygen molecule.

5.2 The current measurement range is 0–140 mA;

5.3 Detection accuracy ≤ ±0.001 mA;

6. Software control system:

6.1 Software control system with independent intellectual property rights;

6.2 The software collects data and can generate real-time data curves.

6.3 Hosting System: Professional editions of operating systems such as Windows 7/8/10;

6.4 Data storage method: Data can be stored in real time during the experiment;

6.5 Equipped with a cloud platform for remote data access;

V. Standard Configuration

Serial number

Name

Quantity

Unit

Note

1

Host

1

Taiwan

 

2

Reaction flask

24

one

 

3

Electrolysis bottle

24

one

 

4

Air cylinder

24

one

 

5

U-shaped tube

24

one

 

6

Connecting hose

1

set

 

7

Software

1

set

 

8

Random file

1

set

 

Electrolytic Respiration-Based Biodegradation Tester