Product


Compost Weighing Method Biodegradability Tester

Compost Weighing Method Biodegradability Tester

Brand: Thmorgan
Model: M1200W Plus

I. Product Applications

1. Meet the testing requirements of GB/T19277.2 standard;
2. Can be used for biodegradability testing via composting methods;
3. Use the weighing method to determine the amount of carbon dioxide produced;
4. Determination of biodegradability for bio-based products, plastics, and similar materials;

II. Product Features  

1. Determine the carbon dioxide release of the sample using the automatic weighing method;

2. 12-channel detection system for high experimental efficiency;

3. The weight of carbon dioxide can be determined online and displayed in graphical form;

4. Automatic weighing, automatic mixing, automatic water addition, and automatic drainage;

5. Adopt a solid-state culture and open-oxygen supply design approach to simulate a highly aerobic environment.

6. Air bath temperature control with temperature adjustment accuracy up to ±0.1℃;

Three, Product Principle

This experimental system can simulate the biodegradation process of bio-based materials under aerobic composting conditions. In an aerobic environment, microorganisms metabolize the bio-based materials as a carbon source, consuming oxygen and producing inorganic byproducts such as carbon dioxide and water. The generated carbon dioxide is immediately absorbed by the alkaline substance contained in the absorption bottle. After the automatic weighing procedure is initiated, a robotic arm drives a mirror-polished stainless steel platform downward, causing the absorption bottle to naturally rest on the weighing sensor. The sensor then collects weight data and transmits it in real time to the PLC controller. Once the weighing is complete, the robotic arm lifts the absorption bottle upward, completely disengaging it from the weighing sensor. After a preset time interval, the system automatically starts the next round of weighing. By continuously monitoring the weight of the absorption bottle via the automatic weighing unit, the incremental weight gain represents the amount of carbon dioxide released during the sample’s biodegradation. The percentage of the sample’s actual carbon dioxide release relative to its theoretical carbon dioxide release constitutes the biodegradation rate.

IV. Technical Parameters

1. Working conditions:

1.1 Power supply voltage: AC 220V ±10%, 50Hz, single-phase;

1.2 Operating temperature: 15–30℃;

1.3 Relative humidity: ≤80%;

1.4 Durability of instrument operation: Capable of continuous operation for extended periods;

2. Oxygen supply system:

2.1 Pre-carbonization system: Available;

2.2 Gas supply method: Continuous aeration for oxygen supply, with a built-in gas supply pump;

2.3 Air pump control: The air pump has 12 channels, each of which can be adjusted independently;

2.4 Flow control method: PLC-controlled; the flow rate can be automatically adjusted according to the set value.

2.5 Flow control accuracy: ±1%;

2.6 Flow Measurement Method: Real-time collection using an electronic flowmeter;

2.7 Flow rate measurement range: 0–300 ml/min;

2.8 Equipped with 12 electronic flow meters for flow rate acquisition;

2.9 Traffic control and traffic collection operate independently;

3. Temperature control system:  

3.1 Temperature control method: Air bath;

3.2  Set temperature range: room temperature – 90℃, with a step size of 0.1℃;

3.3 Temperature control accuracy: ≤ ±0.1℃;

4. Reactor System:

4.1 Number of channels: ≥12;

4.2 Reactor volume: ≥2000 ml;

4.3 Reactor Design: Double-layer design;

4.4 Water-filling method: Automatic water filling;

4.5 Mixing method: Automatic mixing;

4.6 Agitator Shaft Design: Bottom Agitation;

4.7 Gas supply method: Bottom gas supply;

5. Condensation dehumidification system:

5.1 System Components: Low-temperature constant-temperature bath, fully automatic condensation bottle, and dehumidification bottle;

5.2 Channel Configuration: Each channel is equipped with a condensation bottle and a dehumidification bottle.

5.3 Condensation Method: The condenser bottle uses water bath condensation.

5.4 Condensing Temperature: The condensing temperature can be as low as -4℃;

5.5 Drainage Method: Condensate can be automatically discharged at scheduled intervals;

5.6 Dehumidification Method: The dehumidifying bottle features a double-layer design;

5.7 Connection Method: The condenser and the dehumidifier are designed in a series configuration;

6.1 Number of channels: 12 channels;

6.2 Inlet gas method for the absorption bottle: bottom supply;

6.3 Number of absorption bottles: 2 per channel, totaling 24;

6.4 Absorption bottle connection method: In series;

6.5 Absorption Bottle Design: Double-layer design;

6.6 Absorption bottle capacity: 500 mL;

7. Automatic weighing unit:

7.1 Weighing method: Automatic weighing;

7.2 Weighing Principle: Weighing sensor detection, PLC data processing;

7.3  Weighing range: 0-2000g;

7.4  Weighing accuracy: ±0.01 g;

7.5  Weighing interval: 0–48 hours, freely adjustable;

7.6 Continuous operating time: Can be used continuously for long periods;

7.7 Water removal system: Available;

8. Software System:

8.1  Software control system with independent intellectual property rights;

8.2 The real-time page displays the flow rate of each channel and the cumulative weight of carbon dioxide.

8.3 Corresponding cumulative curves and flow rate curves can be generated based on the collected data; the curves can be updated in real time.

8.4 The weighing time interval can be set automatically;

8.5 Data for any time period can be exported;

8.6 Hosting System: Professional editions of operating systems such as Windows 7/8/10;

8.7 Data Storage Method: Data can be stored in real time during the experiment;

8.8 Equipped with a cloud platform for remote data access.

 

Five, Standard configuration

Serial number

Name

Quantity

Unit

Note

1

Host

1

Taiwan

 

2

Reaction flask

12

one

 

3

Decarbonization unit

1

set

 

4

Condensation flask

12

one

 

5

Dehumidifying bottle

12

one

 

6

Low-temperature constant temperature bath

1

set

 

7

Absorption bottle

24

one

 

8

Connecting hose

1

set

 

9

Random file

1

set

 

 

Compost Weighing Method Biodegradability Tester