  一、product uses:

  1. It can be used for the detection of biodegradation by activated sludge method, meeting the requirements of GB / 19276-1 and GB / T 19276-2;

  2. It can be used for soil biodegradation detection, meeting the requirements of GB / T 22047-2008;

  3. It can be used to detect the biodegradability of plastics in seawater / sandy sediment, ISO 18830;

  4. Meet the international standards ISO 14851, ISO 14852, ISO 17556;

  5. Rapid biodegradability test of chemicals, GB / T 21801, GB / T 21802, GB / T 21803;

  6. Meet the requirements of oecd301c, oecd301f, oecd302c, ec-partc4, is05815-1:2003;

  7. Meet the requirements of OECD, EEC, ASTM, BBA, ISO, MITI test research;

  8. It can be used for the determination of bod3, BOD5, bod7, bod28, bod42, etc;

  9. Microbial respiration test;

  10. Microbial breeding experiment;

  二、Product features:

  1. High flux with 12 detection channels;

  2. Oxygen consumption and CO2 production were detected simultaneously to verify each other, so as to improve the accuracy of the results;

  3. Continuous dynamic oxygen supply can be carried out without diluting the sample;

  4. Temperature control system: 5-50 ℃ can be set, and the temperature adjustment accuracy is ± 0.5 ℃;

  5. Air bath heating, no corrosion and electric leakage;

  6. The measurement accuracy is high, which can capture a small change of respiratory rate;

  7. The detection time is humanized, and the exact oxygen consumption curve can be displayed;

  8. Vertical design, compact structure and small space occupation;

  三、Working principle:

  Biological oxygen consumption to determine the biodegradation rate: four charge movements in the electrolysis system produce one molecule of oxygen. The oxygen production is determined by the charge movement. The ratio of actual oxygen consumption to theoretical oxygen consumption is the biodegradation rate of the tested sample.

  Determination of biodegradation rate by carbon dioxide production: carbon dioxide produced by biodegradation reaction is absorbed by sodium hydroxide solution. After absorbing carbon dioxide, the conductivity of the solution changes. The production of carbon dioxide is determined by the change of conductivity. The ratio of actual production of carbon dioxide to theoretical production is the biodegradation rate of the tested sample.

  The two kinds of test results verify each other, so as to ensure that the test results are more accurate.

  四、Technical parameters:

  1. Continuous dynamic oxygen supply system:

  1.1 channels: 12;

  1.2 oxygen supply mode: oxygen is produced by electrolysis of copper sulfate solution;

  1.3 internal solution of electrolytic bottle: copper sulfate solution;

  1.4 cell capacity: 500ml;

  1.5 electrolytic bottle material: transparent hard glass;

  1.6 electrode materials: platinum (cathode), copper (cathode);

  1.7 electrode material of pressure gauge: stainless steel;

  1.8 culture flask capacity: 500ml;

  1.9 culture bottle material: transparent hard glass;

  2. Constant temperature culture system:

  2.1 temperature control mode: air bath;

  2.2 set temperature range: 5 ℃ - 50 ℃;

  2.3 temperature accuracy: ± 0.5 ℃;

  2.4 display screen: LCD;

  2.5 safety door: coating insulation ultraviolet absorption safety door;

  2.6 power supply: 230V AC, 50-60Hz;

  2.7 power fuse: 4am;

  2.8 alarm prompt: Yes;

  2.9 external dimension (WHD): 802mm × 1940mm × 650mm;

  3. Power mixing system:

  3.1 mixing mode: clockwise mixing;

  3.2 drive: 220 / 230 V / 50 / 60 Hz / 32 W;

  3.3 speed: 50-500rpm

  3.4 number of mixing points in each layer: 6;

  4. Detection system:

  4.1 the movement of charge is detected by coulometer to measure oxygen consumption;

  4.2 the carbon dioxide production was measured by conductivity meter;

  4.3 oxygen production rate: 0.25mg/30sec;

  4.4 communication: PC communication via USB technology;

  5. Software system:

  5.1 data acquisition and processing can be realized;

  5.2 adopt the operation mode of menu and window;

  5.3 control with mouse and / or keyboard;

  5.4 computer configuration: dual core, 160g, brand computer.

  5.5 bearing system: any system from Windows2000 to win10;

  5.6 data storage format: Excel or quattropro, etc;

  5.7 data display mode: table or graph;

  5.8 support printing No: Yes;.

  五、Standard configuration:

  1. A set of constant temperature culture system;

  2. 12 sets of 500ml culture bottles;

  3. 12 sets of 500ml electrolytic bottles;

  4. 12 sets of pressure gauges;

  5. A set of software;

  6. One brand computer.

We could not find any corresponding parameters, please add them to the properties table
Previous
Next
托摩根

Focus on us

Beijing Thmorgan Biotechnology Co., Ltd

Telephone:4000-688-151

Mailbox:info@thmorgan.com

Address:2711, building 2, Beijing big data intelligent industrial park, 6 Liye Road, Huilongguan town, Changping District, Beijing

© 2020-2030. All Rights Reserved By Thmorgan.

备案号:京ICP备:12046917号   

Terms of use   Privacy statement   Website map